The specimen orientation allows for simple po-

sition correction of the specimen surface when

the specimen is clamped into place.
• Run the specimen to the rear end position

by turning the coarse feed wheel (see Fig. 6,

page 11). (see

Chapter 5.6

).

• Loosen the clamping lever in front on the mi-

crotome base plate and slide the knife hold-

er base with knife holder until it is almost in

front of the specimen.

For additional information, refer to

Fig. 10,

p. 18

or

Fig. 27,p. 31

.

Specimen blocks must NOT be orient-

ed during the retraction phase!

If a block is oriented during retrac-

tion, the block will advance by the

retraction value PLUS the selected

section thickness before the next

section.

This may cause damage to both speci-

men and knife!

5.5 Orienting the specimen

(only a directional specimen holder fixture)

• Move the specimen holder fixture to the up-

per end position by turning the handwheel

and engage the handwheel lock.

• To release the clamp, turn the eccentric

lever (29) counterclockwise.

• Turn setscrew (31) to orient the specimen in

north-south direction. Turn setscrew (30) to

orient the specimen in east-west direction.

• To lock the current orientation, turn the

eccentric lever (29) clockwise.