Instructions for Use, V2.1 – 10/2012

6.2 Maintenance instructions

Only authorized and qualified service personnel may access the internal components of the

instrument for service and repair!

The instrument is basically maintenance-free. To ensure trouble-free

operation of the instrument over a long period of time, the following is

recommended:

6. Cleaning and Maintenance

• Have the instrument checked at least once a

year by a qualified service technician autho-

rized by Leica.

• At the end of the warranty period, enter into

a service contract. For more information,

please contact your local Leica technical

service center.

• Clean the instrument every day.
• Once a month, lubricate the following parts

with the included oil No. 405 (1-2 drops are

enough):

- The moveable parts of the specimen hold-

er fixture (10) and the dovetail receptacle

(Fig. 14).

- T-piece (19) on the microtome base plate.

- Clamping levers (2) and (20) on the micro-

tome.

- Guide rails (91) for the knife holder base on

the microtome base plate.

- The clamping lever (39) and (24) on the right

and left on the knife holder base.

- The guide (92) of the lateral movement on

the knife holder base (21).

- The T-piece (26) on the knife holder base

(21).

- The sliding surfaces of the knife guard (3)

and knurled nuts (57) on knife holder N.

- The clamping lever (89) on knife holder E.
- Shaft (90) of the clamping lever of the cas-

sette clamp (see Fig. 36).

Fig. 37

14

91

20

2

10

19

24

39

26

92

89

3

57

21