Troubleshooting, 1 possible faults, Possible faults – Leica Biosystems RM2125 RTS User Manual

Page 57

Advertising
background image

57

Leica RM2125 RTS

In the following table there is a list of the most common problems which can arise while

working with the instrument, along with possible causes and troubleshooting procedures.

8. Troubleshooting

Problem

Corrective action

Possible cause

1. Thick/thin sections
The sections alternate between

being thick and thin, or there is

chattering in the sections, or

the specimen is torn out of the

embedding. In extreme cases,

there are no sections whatso-

ever.

• Reclamp the blade, knife holder

or orientation.

• Blade, knife holder or orienta-

tion is not clamped properly.

• Check that the cassette is securely

clamped in the universal cassette
clamp.

• If the universal cassette clamp is

soiled by wax, clean the UCC (refer to
Chapter 8.1 – "Cleaning and mainte-
nance – Universal Cassette Clamp").

• When using cassettes whose lid is

molded on, check that the broken
edge allows the cassette to be
clamped securely; if necessary,
remove burrs or clamp the cassette
into the universal cassette clamp
horizontally instead of vertically.

• If the cassette dimensions are within

the specified tolerances and the cas-
sette still cannot be securely clamped
into place, it may be that the universal
cassette clamp is configured incor-
rectly or is defective. In this case,
have Technical Service inspect and
reconfigure the universal cassette
clamp.

• When using cassettes, particularly

thin-walled cassettes, from a com-
pany other than Leica or Surgipath,
the cassette may become deformed
or other clamping problems may
arise.

If while trying to clamp the cassette

you realize that it is not securely
clamped into place, an alternative
tensioning clamp must be used.

• The specimen is not clamped

properly.

8.1 Possible faults

Advertising