5. Operation

5.9.1 Standard specimen clamp (optional)

The standard specimen clamp is suitable for

holding specimens of 40 x 40 mm.

They are designed for direct clamping of rect-

angular blocks. In addition, they accommodate

the foil clamps.
• Turn the knurled screw (40) counterclock-

wise to move the lower movable jaw (41)

downward.

• Mount the specimen (42) as required.
• Turn the knurled screw (40) clockwise to

move the lower jaw upward against the

fixed jaw to clamp the specimen securely.

5.9 Accessories

All specimen clamps available as accessories can be integrated into either the directional or

non-directional specimen holder fixture.

Fig. 19

40

41

42

When clamping cassettes, make sure

that they are not clamped too tightly,

since the cassette bodies can bend

and result in sections that are

too thick or thin, or the entire speci-

men may fall out and become dam-

aged.