Operation – Leica Biosystems RM2125 RTS User Manual

Page 39

Advertising
background image

39

Leica RM2125 RTS

Adjusting the front pressure plate

• Bring the pressure plate (86)

into position, insert the clamp-

ing lever (89), and use the

clamping lever to clamp the

pressure plate only lightly.

• Adjust the height of the pres-

sure plate using the screws

(87).

The upper edges of the two

pressure plates (86) and (80)

must be at the same height and

parallel to one another.

The clearance angle of the front

pressure plate (86) is adjusted with

a stud screw (88) that is accessible

at an angle from inside through a

bore on the bottom of the knife

holder (

Fig. 33/

).

• Insert a blade (77) and clamp it

lightly using the clamping lever

(89).

The height of the front pressure plate can be adjusted using the stud

screws (87) on the bottom of the knife holder. They can be accessed

through bores on the bottom of the knife holder. An Allen key No. 2 is

required to make the adjustment.

• Use the screw (88) to adjust the pressure plate (86) so that only the up-

per edge of the pressure plate puts pressure on the blade. A gap must

be visible (

Fig. 33/

). A small screwdriver (approx. 3.0 x 70) is required

to make the adjustment.

• When making the adjustment, make sure that the distance between

both pressure plates is approx. 0.4-0.8 mm when open (

Fig. 33/

).

Fig. 33

87

88

86

approx. 0.05 mm

0.4 - 0.8 mm

89

5. Operation

77

Advertising