Operation – Leica Biosystems RM2125 RTS User Manual

Page 38

Advertising
background image

38

Instructions for Use, V2.1 – 10/2012

After each uninstallation or replacement, make sure the pressure

plate sits correctly.

If necessary, readjust it.

The rear pressure plate (80) rests on two stud screws (83,

Fig. 31)

, which

enable height and parallel adjustment.

They can be accessed through bores on the bottom of the knife holder. An

Allen key No. 2 is required to make the adjustment.

5. Operation

Adjusting the rear pressure plate

0.1 - 0.2 mm

77

80

84

85

• Insert the pressure plate (80) and tighten the

screws (82) only far enough that the pres-

sure plate remains moveable.

• Adjust the pressure plate with the stud

screws (83 -

see Fig. 31

) so that the contact

surface (84) for the blade (77) stands approx.

0.1 - 0.2 mm higher than the footprint of the

side flanges of the knife holder (85).

This is important primarily for instruments

whose knife holder base has no lateral

movement.

• When making an adjustment, make sure that

the pressure plate is aligned parallel to the

side flanges of the knife holder.

• Tighten the screws (82).

Fig. 32

Advertising