Viewing & resetting min/max readings – RKI Instruments 12 VDC Beacon 200 User Manual

Page 31

Advertising
background image

12 VDC Beacon 200 Operator’s Manual • 27

the unit may instead shut off if the input voltage decreases below 10.8 V.

Viewing & Resetting Min/Max
Readings

The Reset switch may be used to view and reset the minimum and
maximum gas readings for the active channel(s).

1. While the 12 VDC Beacon 200 is in normal operation, press and hold

the Reset switch button for 3 seconds.

2. The display will indicate MIN / MAX Display Press RESET when

done viewing . . . for 5 seconds before displaying the minimum and
maximum readings for the active channel(s). The minimum reading is
on the left and the maximum is on the right side of the display for each
channel.

3. Press and release the Reset switch button to exit the min/max screen.

The display will indicate To RESET Min/MAX values Press and
HOLD RESET Button
for 10 seconds and then return to normal
operation.

To return to normal operation without resetting the minimum and
maximum readings, do not press the Reset switch button and allow
the unit to return to normal operation.

To reset the minimum and maximum readings, before the unit
returns to normal operation press and hold the Reset switch button
until the display indicates Min/Max Values Have Been Reset.
Release the Reset switch button.The unit will then return to normal
operation.