Chapter 6: input setup program, Overview – RKI Instruments 12 VDC Beacon 200 User Manual

Page 37

Advertising
background image

12 VDC Beacon 200 Operator’s Manual • 33

Chapter 6: Input Setup Program

Overview

This chapter describes how to use the Input Setup Program to add a
channel or change the channel type of an installed channel on the 12 VDC
Beacon 200. The Input Setup Program allows you to define the type of
detector head, the units and gas type, and the full scale for that channel.

To enter the Input Setup Program, the 12 VDC Beacon 200 must first be
off. While the 12 VDC Beacon 200 is off, press and hold the ENTER button,
then turn on the 12 VDC Beacon 200 with the ON/OFF switch.

The Input Setup Program menu includes a 5-minute time-out feature. If you
do not press a button for 5 minutes, the 12 VDC Beacon 200 automatically
begins normal operation.

NOTE: If the 12 VDC Beacon 200 enters normal operation because of a

program time-out, the active channels enter a warm-up period just
as they do when the unit is first turned on.

Setting Up a New Channel or
Changing an Existing Channel

1. While the 12 VDC Beacon 200 is off, press and hold the ENTER

button, then turn on the 12 VDC Beacon 200 with the ON/OFF switch.

2. The 12 VDC Beacon 200 will beep repeatedly while you are holding

down the ENTER button and then the screen will show INPUT SETUP
PROGRAM
on the top line.

3. Press the ENTER button to continue.

If you press the ESCAPE button, the unit will start-up and enter its
warm-up period.

4. Press the UP/YES or DOWN/NO button until the channel you wish to

add or change appears on the display screen, then press the ENTER
button. The detector head type screen appears.

5. The display screen indicates what type of detector head is currently

selected for that channel and asks if it is OK. Table 1 below briefly

Advertising