RKI Instruments 12 VDC Beacon 200 User Manual

Page 45

Advertising
background image

12 VDC Beacon 200 Operator’s Manual • 41

Entering the Calibration Program

1. Assemble the calibration kit(s). See the instruction manual for each

detector head for procedures specific to that detector head.

2. Simultaneously press and hold the ENTER and UP/YES buttons for

approximately 5 seconds to enter the Calibration Program. Release the
buttons when the CALIBRATION PROGRAM... message displays and
asks if you want to continue or return to normal operation.

NOTE: While in the Calibration Program, the alarm status of the 12 VDC

Beacon 200 will be locked in the state it was when the Calibration
Program was entered.

3. Press the ENTER button to continue and display the Calibration Time-

out setting.

The 12 VDC Beacon 200 will remain in the Calibration Program for the
amount of time indicated by the Calibration Time-out setting or until
you exit the program. If necessary, adjust the setting using the UP/YES
and DOWN/NO buttons. Make sure you have allotted enough time to
perform the calibration procedure. Consider the type of detector
head(s) installed and the distance from the 12 VDC Beacon 200 when
determining the time required.

NOTE: If you do not accept a Time-out setting and stay at this screen or

press ESCAPE and do not press control buttons again, then the
12 VDC Beacon 200 will exit the Calibration Program and begin its
warm-up period 5 minutes after the last button was pushed. This
warm-up period functions as if the unit were just powered up.

4. Press the ENTER button to accept the Time-out setting, start the Time-

out period and continue in the Calibration Program.

NOTE: The following describes calibration with a typical channel

allocation which may be different from yours. This example
allocation has two active channels with both types of detector
heads installed. Channel 1 is a 4-20 mA detector head and
channel 2 is a direct connect detector head. See the detector
head instruction manuals for calibration information and
procedures specific to the detector heads.

Advertising