RKI Instruments 12 VDC Beacon 200 User Manual

Page 38

Advertising
background image

34 • 12 VDC Beacon 200 Operator’s Manual

describes each type.

If the detector head type is not correct, press the DOWN/NO button
and continue with step 6.

If the detector head type is correct, press the UP/YES button to
proceed to the units and gas type screen and skip to step 7.

NOTE: See the detector head operator’s manual and the Beacon 200

Detector Head Specification sheet for the detector head to
determine the detector head type.

6. Use the UP/YES and DOWN/NO buttons to scroll through the list of

detector head types until the correct one is displayed. Press the
ENTER button to accept the type.The units and gas type screen
appears.

7. If the units and gas type are not correct, press the DOWN/NO button

and continue with step 8.

If the units and gas type are correct, press the UP/YES button to
proceed to the full scale screen and skip to step 9.

8. Use the UP/YES and DOWN/NO buttons to scroll through a list of units

and gas type choices.

Table 10: 12 VDC Beacon 200 Detector Head Types

Detector Head Type

Description

4 - 20 mA Amp [+S(-)]

A 4 - 20 mA detector head is connected to the 12 VDC
Beacon 200 with 2 or 3 wires, depending on the detector
head model, using the AMP/PREAMP +, S, and - terminals
on a detector head terminal strip. All calibration adjustments
are made at the detector head.

PreAmp [+S]

A PreAmp detector head is connected to the 12 VDC Beacon
200 with 2 wires using the AMP/PREAMP + and S terminals
on a detector head terminal strip. All calibration adjustments
are made at the 12 VDC Beacon 200.

O2 Direct [+-]

An O2 Direct detector head is an oxygen detector head in
which the oxygen detector is wired to the 12 VDC Beacon
200 with 2 wires using the OXYGEN + and - terminals on a
detector head terminal strip. All calibration adjustments are
made at the 12 VDC Beacon 200.

LEL Direct [RWGB]

An LEL Direct detector head is a combustible gas detector
head in which the combustible gas detector is wired to the 12
VDC Beacon 200 with 4 wires using the LEL R, W, G, and B
terminals on a detector head terminal strip. All calibration
adjustments are made at the 12 VDC Beacon 200.

Advertising