RKI Instruments 12 VDC Beacon 200 User Manual

Page 35

Advertising
background image

12 VDC Beacon 200 Operator’s Manual • 31

10. Press the UP/YES button to save the configuration. The screen will

then return to the Channel Control & Setup menu.

11. To view or change the Channel 2 settings, scroll to the Configure

Channels Menu and repeat steps 4 through 10.

12. To exit the Channel Control & Setup menu, press ESCAPE to return to

the screen which asks Control & Setup Program Proceed? [YES] or
[NO]
.

13. Press the DOWN/NO button to return to normal operation.

Table 9: Channel Setting Parameters

Parameter

(Factory-Set Value)

Description

ALARM-1 Level
(See the Beacon 200 Detector Head
Specification Sheet for the detector
head installed on this channel)

The gas reading at which the 12 VDC Beacon 200 initiates an alarm 1
condition for this channel.

ALARM-1 ON DELAY
(1 sec)

The amount of time the 12 VDC Beacon 200 delays activation of the
alarm 1 circuit once an alarm 1 condition is initiated.

ALARM-1 OFF DELAY
(0 sec)

The amount of time the 12 VDC Beacon 200 delays turning off the alarm
1 circuit once an alarm 1 condition passes.

ALARM-1 (activation)
(DECREASING for oxygen channels,
INCREASING for all other channel
types)

Indicates if the alarm 1 circuit is activated by gas readings INCREASING
or DECREASING to the ALARM-1 Level.

ALARM-1 Relay (action)
(NORMALLY DE-ENERGIZED)

If set as NORMALLY DE-ENERGIZED, the channel’s alarm 1 relay is de-
energized in normal operation and energizes when an alarm 1 condition
is initiated.
If set as NORMALLY ENERGIZED, the channel’s alarm 1 relay is
energized in normal operation and de-energizes when an alarm 1
condition is initiated.

ALARM-1 Relay (reset)
(SELF RESETTING for oxygen,
LATCHING for all other channel types)

If set as LATCHING, you must press the RESET button to reset the
alarm 1 circuit after the alarm 1 condition passes.
If set as SELF RESETTING, the 12 VDC Beacon 200 automatically
resets the alarm 1 circuit after the alarm 1 condition passes.

ALARM-2 Relay (used for)
(ALARM-2 Condition)

If set as ALARM-2 Condition, the channel’s alarm 2 relay activates
when an alarm 2 condition is initiated for the channel.
If set as FAIL Condition, the channel’s alarm 2 relay activates when a
fail condition is initiated for the channel.

ALARM-2 Level
(See the Beacon 200 Detector Head
Specification Sheet for the detector
head installed on this channel)

The gas reading at which the 12 VDC Beacon 200 initiates an alarm 2
condition for this channel.

ALARM-2 ON DELAY
(2 min for oxygen, 1 sec for all other
channel types)

The amount of time the 12 VDC Beacon 200 delays activation of the
alarm 2 circuit once an alarm 2 condition is initiated.

Advertising