Figure 11. direct connect detector head(s) only – RKI Instruments 12 VDC Beacon 200 User Manual
Page 42
Advertising

38 • 12 VDC Beacon 200 Operator’s Manual
to guide you through the program. At any point in the calibration program,
the ESCAPE key may be used to either return to the previous screen or
abort a process.
Select Cal. Gas
Concentration
for Channel
2
Calibrate
Channel 1 Y/N?
Calibration
Timeout
Selection
ENTER
Normal
Operation
Apply
Gas to
Ch. 1 & Ch. 2
Detectors
Air Adjust
Channel
2
(If Installed)
Press Enter to
Adjust Span
Span
Channel 2 Y/N?
(If Installed)
Select Cal. Gas
Concentration
for Channel
1
Air Adjust
Channel
1
Calibration
Program Enter
ENTER/ESCAPE
ESCAPE
Calibrate
Channel 2 Y/N?
(If Installed)
Span
Channel 1 Y/N?
Figure 11. Direct Connect Detector Head(s) Only