Beacon 200 Rig Monitor Operator’s Manual • 30

10. Press the UP/YES button to save the configuration. The screen will

then return to the Channel Control & Setup menu.

11. To view or change the Channel 2 settings, scroll to the Configure

Channels Menu and repeat steps 4 through 10.

12. To exit the Channel Control & Setup menu, press ESCAPE to return to

the screen which asks Control & Setup Program Proceed? [YES] or
[NO]
.

13. Press the DOWN/NO button to return to normal operation.

Table 8: Channel Setting Parameters

Parameter

(Factory-Set Value)

Description

ALARM-1 Level
(10 ppm)

The gas reading at which the Rig Monitor initiates an alarm 1 condition
for this channel.

ALARM-1 ON DELAY
(1 sec)

The amount of time the Beacon 200 delays activation of the alarm 1
circuit once an alarm 1 condition is initiated.

ALARM-1 OFF DELAY
(0 sec)

The amount of time the Beacon 200 delays turning off the alarm 1 circuit
once an alarm 1 condition passes.

ALARM-1 (activation)
(INCREASING)

Indicates if the alarm 1 circuit is activated by gas readings INCREASING
or DECREASING to the ALARM-1 Level.

Note: Changing this parameter will affect the operation of the factory
installed strobe/horn. Do not change this parameter from the factory
setting.

ALARM-1 Relay (action)
(NORMALLY DE-ENERGIZED)

If set as NORMALLY DE-ENERGIZED, the channel’s alarm 1 relay is de-
energized in normal operation and energizes when an alarm 1 condition
is initiated.
If set as NORMALLY ENERGIZED, the channel’s alarm 1 relay is
energized in normal operation and de-energizes when an alarm 1
condition is initiated.

Note: Changing this parameter will affect the operation of the factory
installed strobe/horn. Do not change this parameter from the factory
setting.

ALARM-1 Relay (reset)
(LATCHING)

If set as LATCHING, you must press the RESET button to reset the
alarm 1 circuit after the alarm 1 condition passes.
If set as SELF RESETTING, the Rig Monitor automatically resets the
alarm 1 circuit after the alarm 1 condition passes.

ALARM-2 Relay (used for)
(ALARM-2 Condition)

If set as ALARM-2 Condition, the channel’s alarm 2 relay activates
when an alarm 2 condition is initiated for the channel.
If set as FAIL Condition, the channel’s alarm 2 relay activates when a
fail condition is initiated for the channel.

ALARM-2 Level
(50 ppm)

The gas reading at which the Rig Monitor initiates an alarm 2 condition
for this channel.

ALARM-2 ON DELAY
(1 sec)

The amount of time the Rig Monitor delays activation of the alarm 2
circuit once an alarm 2 condition is initiated.