RKI Instruments Beacon 200 RIg Monitor, H2S/H2S User Manual
Page 50

45 • Beacon 200 Rig Monitor Operator’s Manual
Slow or No
Response/
Difficult or
Unable to
Calibrate
•
Rig Monitor responds
slowly or does not
respond to response
test.
•
Unable to successfully
set the fresh air or
span reading during
calibration.
•
Rig Monitor requires
frequent calibration.
Note:
Under “normal”
circumstances, the
detector requires
calibration once every 3
months.
Some applications may
require a more frequent
calibration schedule.
•
The calibration cylinder
is low, out-dated, or
defective.
•
The flame arrestor in the
detector housing cap is
wet or clogged with dirt
or other particulates.
•
The detector is
malfunctioning.
•
The main PCB is
malfunctioning.
1.
Verify that the calibration cylinder contains
an adequate supply of a fresh test sample.
2.
Verify that the calibration gas
concentration is appropriate for the
detector. Zero emission air (0 ppm H
2
S) is
normally used for a zero adjustment if the
environment is suspect and 25 PPM H
2
S
in nitrogen is normally used for a response
adjustment.
3.
Check the detector housing cap to
determine if the flame arrestor is wet or
dirty. Clean if necessary.
4.
If the calibration/response difficulties
continue, replace the detector.
5.
If the calibration/response difficulties
continue, contact RKI for further
instruction.
Table 9: Troubleshooting the Beacon 200 Rig Monitor (Continued)
Condition
Symptom(s)
Probable Causes
Recommended Action