RKI Instruments Beacon 200 RIg Monitor, H2S/H2S User Manual

Page 50

Advertising
background image

45 • Beacon 200 Rig Monitor Operator’s Manual

Slow or No
Response/
Difficult or
Unable to
Calibrate

Rig Monitor responds
slowly or does not
respond to response
test.

Unable to successfully
set the fresh air or
span reading during
calibration.

Rig Monitor requires
frequent calibration.

Note:

Under “normal”

circumstances, the
detector requires
calibration once every 3
months.

Some applications may

require a more frequent
calibration schedule.

The calibration cylinder
is low, out-dated, or
defective.

The flame arrestor in the
detector housing cap is
wet or clogged with dirt
or other particulates.

The detector is
malfunctioning.

The main PCB is
malfunctioning.

1.

Verify that the calibration cylinder contains
an adequate supply of a fresh test sample.

2.

Verify that the calibration gas
concentration is appropriate for the
detector. Zero emission air (0 ppm H

2

S) is

normally used for a zero adjustment if the
environment is suspect and 25 PPM H

2

S

in nitrogen is normally used for a response
adjustment.

3.

Check the detector housing cap to
determine if the flame arrestor is wet or
dirty. Clean if necessary.

4.

If the calibration/response difficulties
continue, replace the detector.

5.

If the calibration/response difficulties
continue, contact RKI for further
instruction.

Table 9: Troubleshooting the Beacon 200 Rig Monitor (Continued)

Condition

Symptom(s)

Probable Causes

Recommended Action

Advertising