33 • Beacon 200 Rig Monitor Operator’s Manual

NOTE: The following procedure assumes that the target gas is present in

a high enough concentration to affect the fresh air (zero) reading.
If a fresh air environment can be verified, applying zero air to the
detector is not necessary when setting the zero reading.

This section describes how to calibrate the H

2

S detectors. It includes

procedures to prepare for calibration, set the fresh air reading, set the span
reading, and return to normal operation. It describes calibration using a
calibration kit that includes calibration gas, sample tubing, and a fixed flow
regulator with an on/off knob. RKI Instruments, Inc. recommends using a
0.5 LPM (liters per minute) fixed flow regulator. The calibration cup/splash
guard is part of the detector assembly in the Rig Monitor.

Preparing for Calibration

1. Simultaneously press and hold the ENTER and UP/YES buttons for

approximately 5 seconds to enter the Calibration Program. Release the
buttons when the CALIBRATION PROGRAM... message displays and
asks if you want to continue or return to normal operation.

Select Cal. Gas

Concentration

for Channel

2

Calibrate

Channel 1 Y/N?

Calibration

Timeout

Selection

ENTER

Normal

Operation

Apply

Gas to

Ch. 1 & Ch. 2

Detectors

Air Adjust

Channel

2

(If Installed)

Press Enter to

Adjust Span

Span

Channel 2 Y/N?

(If Installed)

Select Cal. Gas

Concentration

for Channel

1

Air Adjust

Channel

1

Calibration

Program Enter

ENTER/ESCAPE

ESCAPE

Calibrate

Channel 2 Y/N?

(If Installed)

Span

Channel 1 Y/N?

Figure 11. Calibration Flow Chart