9

M2 Transmitter Operator’s Manual

Figure 7: 65-2495RK H

2

S Detector Non-Explosion Proof

ESM-01 Detectors

The construction of the various ESM-01 detectors is similar. Each of the ESM-01 detectors
has a different replaceable plug-in sensor that protrudes through the detector housing
cap. Although ESM-01 detectors look the same except for the plug-in detector, the detector
housing which includes the detector socket and conditioning circuitry is specific to a
sensor type. For example, a detector housing from a Cl

2

detector cannot be used with an

NH

3

detector. One of the two detector wires is color coded for each specific detector type.

The ESM-01 detectors include a removable splashguard that screws onto the detector
housing cap. When it is necessary to apply gas to an ESM-01 detector, remove the
splashguard to allow access to the sensor that protrudes through the detector housing cap.
You can then push the appropriate ESM-01 calibration cup onto the sensor.

Figure 8: ESM-01 Type Detector

Cap Gasket

Detector Housing Body

Hydrophobic
Membrane

H2S Plug-in Sensor

Detector
Housing Cap

Detector Housing Body
(different for each detector type)

Cap Gasket

Splash Guard
(removable)

Detector Housing
Cap

Sensor
(different for each detector type)

Advertising