M2 Transmitter Operator’s Manual

10. Make controller, device, and relay connections as appropriate. If shielded cable is

used for the PWR/SIG connections, connect the cable shield’s drain wire to an
available chassis ground at the gas monitoring controller, recording device, or
programmable controller.

Start Up

Introducing Incoming Power

1.

Complete the installation procedures described earlier in this manual.

2.

Verify that all wiring connections are correct and secure.

3.

Turn on the incoming power.

4.

If necessary, turn on the controller or other monitoring device that is connected to the
M2.

5.

The LCD display will indicate the firmware version when the M2 is first powered up
and will then count down a one minute warm-up period before normal operation
begins. During normal operation, the display will indicate the target gas and current
gas reading. Verify that the display is indicating the target gas and current gas reading
after the warm-up period is complete and normal operation begins.

NOTE:

Allow the M2’s detector to warm up for 15 minutes before you continue with
the next section, “Adjusting the Fresh Air Reading.”

Adjusting the Fresh Air Reading

When the M2 is shipped from RKI Instruments, Inc., it is factory calibrated. If a full
calibration is desired at startup, see “Calibration, Combustible Gas, CO2, and Toxic
Versions” on page 40 or “Calibration, Oxygen Version” on page 43.

Verify that the M2 is in a fresh air environment (environment known to be free of the
target gas and combustible or toxic gas vapors and of normal oxygen content, 20.9%).

CAUTION:

If you suspect the presence of combustible gas, CO

2

, toxic gas, or that the oxygen

content is not normal in the monitoring environment, use the calibration kit and
the zero air calibration cylinder to introduce “fresh air” to the detector and verify
an accurate fresh air setting. See the “Calibration, Combustible Gas, CO2, and
Toxic Versions” on page 40 or “Calibration, Oxygen Version” on page 43 for
instructions on how to use a zero air cylinder when performing a fresh air
adjustment.

Non-Intrusive Fresh Air Signal Adjustment

If the M2 is installed in a classified area and non-intrusive zero adjustment is required,
follow the instructions below, but do not remove the junction box cover. Use the
magnetic wand accessory to actuate the magnetic switches instead of pushing the control
switch buttons. To actuate a magnetic switch and perform the same operation as pressing
a control switch button, touch the magnet in the end of the magnetic wand to the M2’s
glass cover directly over the magnetic switch you wish to actuate. Touching the glass and

METHANE

0 %LEL