Specifications – RKI Instruments M2 Series User Manual

Page 7

Advertising
background image

M2 Transmitter Operator’s Manual

2

Specifications

Table 1 lists specifications for the M2.

Table 1: M2 Specifications

Target Gas/Detection
Range

Combustible Gas/Carbon Dioxide (CO

2

)

• Catalytic LEL: 0 - 100% LEL, 1% LEL increments (CH

4

calibration

unless otherwise specified. H

2

specific detector available)

• IR (infrared) CH

4

LEL: 0 - 100% LEL, 1% LEL increments

• IR HC LEL: 0 - 100% LEL, 1% LEL increments (propane calibration)
• IR CH

4

Volume: 0 - 100 % volume, 1% increments

• IR CO

2

ppm: 0 - 5,000 ppm, 20 ppm increments

• IR CO

2

Volume: 0 - 5.00 % volume, 0.01% increments

• IR CO

2

Volume: 0 - 50.0 % volume, 0.1% increments

• IR CO

2

Volume: 0 - 100 % volume, 1% increments

Oxygen
• Oxygen (O

2

): 0 - 25.0 % volume, 0.1% increments

CO/H

2

S

• Carbon Monoxide (CO): 0 - 300 ppm, 1 ppm increments
• Hydrogen Sulfide (H

2

S): 0 - 100 ppm, 1 ppm increments

ESM-01 Toxic
• Ammonia (NH

3

): 0 - 75.0 ppm, 0.1 ppm increments

• Arsine (AsH

3

): 0 - 1.50 ppm, 0.01 ppm increments

• Chlorine (Cl

2

): 0 - 3.00 ppm, 0.01 ppm increments

• Chlorine (Cl

2

): 0 - 10.0 ppm, 0.1 ppm increments

• Hydrogen Cyanide (HCN): 0 - 15.0 ppm, 0.1 ppm increments
• Phosphine (PH

3

): 0 - 1.00 ppm, 0.01 ppm increments

• Sulphur Dioxide (SO

2

): 0 - 6.00 ppm, 0.01 ppm increments

Alarm Settings
(Alarm 1/Alarm 2)

Combustible Gas (%LEL):

10 %LEL/50 %LEL

Combustible Gas (%Volume):

100 %/100 %

CO2 0 - 5,000 ppm:

2,500 ppm/5,000 ppm

CO2 0 - 5.00%:

0.50 %/3.00 %

CO2 0 - 50.0%:

50.0%/50.0%

CO2 0 - 100%:

100%/100%

Oxygen:

19.5%(decreasing)/23.5%(increasing)

Carbon Monoxide:

25 ppm/50 ppm

Hydrogen Sulfide:

10 ppm/50 ppm

Ammonia:

12 ppm/25 ppm

Arsine:

0.20 ppm/0.50ppm

Chlorine (0 - 3.00 ppm):

0.50 ppm/1.50 ppm

Chlorine (0 - 10.0 ppm):

1.0 ppm/3.0 ppm

Hydrogen Cyanide:

5.0 ppm/10.0 ppm

Phosphine:

0.10 ppm/0.30 ppm

Sulphur Dioxide:

1.00 ppm/2.00 ppm

Construction
(housing)

Explosion-proof Junction Box, NEMA 4

Advertising