M2 Transmitter Operator’s Manual

52

Termination Jumper

The M2 includes a 2-pin termination header (see Figure 9) that is used when the M2 is
used in a Modbus system. Every M2 is supplied with a termination jumper (a jumper
block) installed onto this header. If the M2 is not used in a Modbus system, this jumper
has no function. When the M2 is installed in a Modbus system, this jumper must be
installed in an M2 that is at the end of a Modbus line. Any M2 in a Modbus system that is
not at the end of a line must have the termination jumper removed (see Figure 29 &
Figure 30 below).

Figure 29: Multiple M2s in a Daisy Chain Configuration

Figure 30: Multiple M2s in a Two Branch Configuration

Modbus

Master

M2

Termination

Jumper

Installed

M2

Termination

Jumper

Not Installed

M2

Termination

Jumper

Not Installed

M2

Termination

Jumper

Not Installed

M2

Termination

Jumper

Not Installed

ID = 128

ID = 127

ID = 126

ID = 125

ID = 1

Up to 128 M2s can be connected without a repeater.

RS-485

RS-485

Modbus

Master

M2

Termination

Jumper

Installed

M2

Termination

Jumper

Not Installed

M2

Termination

Jumper

Not Installed

M2

Termination

Jumper

Not Installed

M2

Termination

Jumper

Not Installed

ID = 1

ID = 2

ID = 3

ID = 4

M2

Termination

Jumper

Installed

ID = 128

ID = 127

Up to 128 M2s can be connected without a repeater.

RS-485

RS-485