M2 Rig Monitor Operator’s Manual

NOTE:

You can select normally energized (N. EN) or normally de-energized (N. DE-
EN) alarm 1 and alarm 2 relay settings in Configuration Mode. The following
sections describe the standard factory setting for these relays which is
N. DE-EN.

The fail relay is factory set as N. EN and is not user-adjustable.

Alarm 1 Condition

Alarm 1 Condition Indications

When the gas reading reaches the alarm 1 setpoint, the Rig Monitor senses an alarm 1
condition. The Rig Monitor alerts you to an alarm 1 condition as follows:

The A1 LED turns on.

The gas reading alternates with the ALARM-1 message.

The alarm 1 relay energizes and the strobe/horn activates.

Responding to an Alarm 1 Condition

1.

Follow your established procedure for a low level H

2

S condition.

2.

After the gas reading falls below the alarm 1 setpoint, press the ENTER button to reset
the alarm 1 circuit. Resetting the alarm 1 circuit turns off the A1 LED, resets the LCD,
de-energizes the alarm 1 relay, and turns off the strobe/horn.

NOTE:

If the ENTER button is pressed while the M2 Rig Monitor is in an alarm 1
condition, the A1 LED will flash but all other indications will remain
unchanged.

You cannot de-energize the alarm 1 relay, and consequently the strobe/horn,
until the gas reading falls below the alarm 1 setpoint.

Fail

• Disconnected or misconnected

detector wiring

• Display reading at -10% of full

scale or lower

• Defective components

• F LED is on
FAIL message replaces

gas reading

NOTE: There is a 30 second
delay on the fail condition.

Low Power

DC power source less than 18.5 volts.

• F LED is on
LowPower message and

actual voltage of

incoming DC power

NOTE: Under typical operating conditions, there may be no low power alarm. The 12 VDC
battery that is recommended for operation may fail before the power converter recognizes a low
power condition.

*

1

If the M2 Rig Monitor is in both an alarm 1 and an alarm 2 condition, both alarm LEDs

are on and the display alternates between the gas reading and the ALMS 1&2 message.

Table 3: Visual and Audible Alarm Indications

Condition

Cause

Visual Indication(s)