M2 Rig Monitor Operator’s Manual

40

Register 3

Register 3 is the supply voltage (0.1 volt per count).

Registers 4 - 11

The 16 bytes in these registers contain the same ASCII text information that is displayed
on the Rig Monitor’s LCD (not NULL Terminated).

Register 12

Register 12 is the range (full scale readout).

Register 13

Register 13 is the alarm 1 set point.
The decimal point location is the same as specified in Register 1.

Register 14

Register 14 is the alarm 2 set point.
The decimal point location is the same as specified in Register 1.

Registers 15 - 19

Registers 15 - 19 are the Gas Name ASCII String (NULL Terminated).

Registers 20 - 22

Registers 20 - 22 are the Gas Units String (NULL Terminated).

Register 23

Register 23 is the alarm 1 trigger.
0=Decreasing
1=Increasing

Register 24

Register 24 is the alarm 1 normal relay state.
0=Normally De-Energized
1=Normally Energized

Register 25

Register 25 is the alarm 1 relay reset.
0=Latching
1=Self-Resetting

Register 26

Register 26 is the alarm 1 ON delay.
Integer values are in seconds.

[0]

Alarm 1 Relay Status
0 = Not Energized
1 = Energized