RKI Instruments M2 Rig Monitor User Manual

Page 29

Advertising
background image

M2 Rig Monitor Operator’s Manual

24

4.

When you have scrolled through all the adjustable parameters, SAVE IT? YES/NO
appears on the display.

5.

If you do not wish to save the adjustments and want to exit Configuration Mode,
press and release the DOWN/NO button. The DO OVER? YES/NO message will
display. Press and release the DOWN/NO button. The ABORT? YES/NO message
will display. Press the UP/YES button to return to normal operation.

If you wish to change some of the adjustments made, press and release the DOWN/
NO button. The DO OVER? YES/NO message will display. Press and release the UP/
YES button. The Re-do Config message will display and the M2 Rig Monitor will
return to the first adjustable parameter. Go back to step 3 and continue.

If you wish to save the adjustments made, press and release the UP/YES button.
Config Saved

is indicated on the display for a few seconds and the M2 Rig Monitor

returns to normal operation.

A2 OffDy (alarm 2 off delay)
(0 sec.)

The amount of time the Rig Monitor delays turning off the alarm 2 circuit
once an alarm 2 condition passes. This parameter appears only if the
alarm 2 relay reset setting is set to SELF-RST. It can be set in 1 second
increments from 0 - 60 seconds, in 1 minute increments from 1 - 15
minutes, and in 15 minute increments from 15 - 60 minutes.

ZeroSupp
(2 ppm)

The zero suppression feature helps prevent “jumpy” readings near the
fresh air reading.
For example, if the zero suppression setting for the H

2

S detector is 2.0

ppm, the Rig Monitor will display a reading of 0 ppm for gas readings
from -2 ppm to 2 ppm.

FILTER
(5 secs)

The filter feature helps “smooth out” jumpy or noisy signals from the
detector. You can set the filter from 0 seconds to 60 seconds in 5
seconds increments.
The displayed gas reading is the average reading over the previous time
period defined by the filter setting. So if the filter is set to 5 seconds, the
displayed gas reading is the average over the past 5 seconds.

CAL Time
(15 mins)

The calibration time-out sets the amount of time after the last button push
while in Calibration Mode that the Rig Monitor will wait before returning to
normal operation. If the calibration time-out is set to 15 mins, then Rig
Monitor will return to normal operation automatically 15 minutes after the
last button push.

Auto Zero
(Enabled)

If set to Enabled, the Rig Monitor’s firmware will monitor the zero (fresh
air) reading and make automatic zero adjustments under precisely
specified conditions if it determines that a reading change is due to
normal sensor drift.
If set to Disabled, the Auto Zero function is turned off.

Table 4: Configuration Parameters (Continued)

Parameter

(Factory Set Value)

Description