Alarm strobe/horn, Magnetic wand, S detector, junction box, and cable – RKI Instruments M2 Rig Monitor User Manual
Page 9
![background image](/manuals/571531/9/background.png)
M2 Rig Monitor Operator’s Manual
4
Alarm Strobe/Horn
The Rig Monitor has a red alarm strobe light/horn installed in the top conduit hub. The
Rig Monitor retains its NEMA 4 rating with the strobe installed. Strobe operation is
controlled by the Alarm 1 relay contacts. The relay action settings can be changed in
Configuration Mode.
Magnetic Wand
The magnetic wand is a short plastic rod with a magnet in one end. It is used to actuate
the magnetic control switches on the control PCB while the junction box cover is still
installed so that non-intrusive calibration can be performed.
H
2
S Detector, Junction Box, and Cable
The H
2
S detector consists of a junction box with a mounting bracket, the detector, and the
calibration cup/splash guard.
Figure 2: H
2
S Junction Box Component Location
Detector Cable (25 ft)
To M2
Rubber
Spacers (3X)
Cable Bushing (25 ft)
Terminal Strip
Mounting Bracket
H2S Detector
(Factory Wired)
Calibration Cup/
Splash Guard