RKI Instruments M2 Rig Monitor User Manual

Page 44

Advertising
background image

39

M2 Rig Monitor Operator’s Manual

Table 9: Register 2, Operating State, Alarms & Relays Bit & Field Assignments

Bit(s)

Value

[15:12]

Operating State
0 = Undefined
1 = Initialization
2 = Normal
3 = Warm-up
4 = Fail
5 = Calibration
6 = Post-Calibration
7 = Low Power
8 = Setup

[11]

Gas Type Change
0=Not Changed
1=Changed

[10]

Configuration Change
0=Not Changed
1=Changed

[9]

Calibration Activity Flag
0=No Calibration Activity
1=Calibration Activity Has Occurred

[8]

Fail Status
0 = Fail Not Asserted
1 = Fail Asserted

[7:6]

Alarm 2 Status
0 = No Alarm
1 = Unacknowledged Alarm
2 = Acknowledged Alarm
3 = Unused Code

[5:4]

Alarm 1 Status
0 = No Alarm
1 = Unacknowledged Alarm
2 = Acknowledged Alarm
3 = Unused Code

[3]

Overscale Status
0 = Not Overscale
1 = Overscale

[2]

Fail Relay Status
0 = Not Energized
1 = Energized

[1]

Alarm 2 Relay Status
0 = Not Energized
1 = Energized