RKI Instruments M2 Rig Monitor User Manual

Page 32

Advertising
background image

27

M2 Rig Monitor Operator’s Manual

Strobe/
Horn Not
Working

• The strobe/horn

does not activate
when an alarm 1
condition occurs.

• The strobe/horn is

malfunctioning.

• The strobe/horn

wiring is
disconnected or
misconnected.

1. Verify that the strobe/horn is

wired properly to the alarm
terminal strip.

2. If the strobe/horn difficulties

continue, contact RKI for further

instruction.

Fail
Condition

• Rig Monitor

indicates a fail

condition.

• The detector wiring

to the terminal PCB

is disconnected or

misconnected.

• The plug-in H

2

S

sensor is not

properly plugged

into the sockets in

the detector

housing body.

• The detector fresh

air signal is low

enough to cause a

fail condition.

• The detector is

malfunctioning.

• The terminal or

display PCB is

malfunctioning.

1. Verify that the detector wiring is

correct and secure.

2. Confirm that the plug-in H

2

S

sensor is installed properly.

3. Calibrate the Rig Monitor.
4. If the fail condition continues,

replace the detector.

5. If the fail condition continues,

contact RKI for further

instruction.

Slow or No
Response/
Difficult or
Unable to
Calibrate

• Rig Monitor

responds slowly

or does not

respond to

response test.

• Unable to

successfully set

the fresh air or

span reading

during

calibration.

• Rig Monitor

requires frequent

calibration.

Note:

Under “normal”

circumstances, the
detector requires
calibration once every 3
months.

Some applications may

require a more frequent
calibration schedule.

• The calibration

cylinder is low, out-

dated, or defective.

• The flame arrestor

in the detector

housing cap is wet

or clogged with dirt

or other

particulates.

• The detector is

malfunctioning.

• The terminal PCB is

malfunctioning.

1. Verify that the calibration cylinder

contains an adequate supply of a

fresh test sample.

2. Verify that the calibration gas

concentration is appropriate for

the detector. Zero emission air (0

ppm H

2

S) is normally used for a

zero adjustment if the

environment is suspect and 25

PPM H

2

S in nitrogen is normally

used for a response adjustment.

3. Check the detector housing cap to

determine if the flame arrestor is

wet or dirty. Clean if necessary.

4. If the calibration/response

difficulties continue, replace the

detector.

5. If the calibration/response

difficulties continue, contact RKI

for further instruction.

Table 5:Troubleshooting the M2 Rig Monitor

Condition

Symptom(s)

Probable Causes

Recommended Action

Advertising