RKI Instruments GD-K71D User Manual

Page 19

Advertising
background image

PT2E-1052

18

4−5.Caution at wiring construction











4−6.Applicable cable
4−6−1.Power/Signal cable

[3-wire 4∼20mA version]
Power & signal cable

: Use CVVS 1.25sq. 3-core cable

[4-wire 4∼20mA version]
Power cable

: Use CVV 1.25sq, 2-core cable

Signal

: Use CVVS 1.25sq, 2-core cable








4−6−2.Relay output signal cable

Relay contact cable : Use VCTF 0.75sq., 6-core cable

4−6−3.Specifications of terminal

Terminal specifications

Rated voltage : AC250V

Rated current : 13A

Connectable conditions

Cable

: 0.14mm2∼1.5mm2

Length of naked wires : Approx 6mm

Screw clamp torque : 5∼8kg.

Applicable screw driver : Minus screw driver(width below 3mm)

When use the bar terminal, the following can be used.

Bar terminal :Model AI series(Phoenix contact make)

Terminal lug tool :Model CRIMPFOX UD 6(Phoenix contact make)






CAUTION

Be sure to use the exclusive bar terminal. When used with other make bar terminal
than above, the function of this detector can not be warranted.

CAUTION

・When make wiring construction, take care not to damage the internal electronics circuit.

・When handle the detector unit, put in at horizontal position. If put it vertically, it tends

to fall back and damage the detector.

・For power cable, and signal cable, they should not be laid out with motive power cable

such as motor etc.

・When use twisted cable, take care not to contact a part of core cable with other core

cable.

CAUTION

Size of leading cable to GD-K71D is upto 1.25sq.
When use the cable exceeding 1.25sq., use the junction box etc.

Advertising