24
Zero suppression
● ○ ○ ○
PW/TR ALM1 ALM2 SKIP
During detection mode operation,
below +6% of full scale shows as
“0”.
● ○ ○ ○
PW/TR ALM1 ALM2 SKIP
WARNING
When it shows 「-0.00」, the accurate gas reading is impossible and make
zero adjustment.
● ○ ○ ○
PW/TR ALM1 ALM2 SKIP
Trouble condition
When any of trouble arises on detector,
content of trouble is shown.
(LED display) (Content of trouble)
E−00 System trouble
E−01 Sensor disconnection or not yet connected.
E−04 Zero correction function trouble.
E−05 Low flow
● ○ ○ ○
PW/TR ALM1 ALM2 SKIP
During detection mode operation,
this shows actual reading from above
+6% of full scale.
The right 「-0.00」 is displayed when it
gets down to minus(-)side more than
10% of full scale.