8-4. Trouble Shooting

(1) PW/TR light is not ON. (Power is not supplied.)
<Cause and Treatment>

・Is it OK with power cable connection ?

Connect power cable.

・Is fuse disconnected ?

Search the cause of disconnection and replace it after making counter-measures.

・Is power voltage normal ?

Supply the rated power voltage.

・Is this power switch turned on ?

Make power switch ON.

(2) Abnormal performance
<Cause and Treatment>

・Sudden serge noise can be considered. When recover it, turn off the power and turn

on the power again for re-start of operation. When such takes place often times,
take an appropriate action for noise.

(3) No calibration available
<Cause and Treatment>

・Is the concentration of calibration gas correct ?

Prepare the appropriate calibration gas.

・There is the possibility for decreasing of sensor sensitivity.

The sensor replacement is required.

(4) PW/TR light flashing.
① System Trouble “E-00”
<Cause and Treatment>

・Memory trouble inside of detector.

Contact our nearest agent or Riken Keiki.

② Sensor connection trouble “E-01”
<Cause and Treatment>

・Non-connection of sensor, contact failure of connector and disconnection of sensor cable.

Check

“whether sensor is connected” or “the sensor cable connector is connected to the

sensor or amplifier board.

If can not be recovered, contact our nearest agent or Riken Keiki.

③ Abnormal function for zero correction 「E-04」

・Zero correction was out of available range due to sensor drift.

Contact our nearest agent or RIKEN KEIKI.

④ Low flow trouble

“E-05”

<Cause and Treatment>

・Drop of pump function

Even though the ball of flowmonitor can not be up the red line(lower red line) with FLOW
ADJ/▲▼switch, replace the pump with new one.

・Check that the detector unit is mounted firmly or the fixing screw of detector unit is

fastened firmly.