8-4. Trouble Shooting

(1) PW/TR light is not ON. (Power is not supplied.)
<Cause and Treatment>

・Is it OK with power cable connection ?

Connect power cable.

・Is fuse disconnected ?

Search the cause of disconnection and replace it after making counter-measures.

・Is power voltage normal ?

Supply the rated power voltage.

・Is this power switch turned on ?

Make power switch ON.

(2) Abnormal performance
<Cause and Treatment>

・Sudden serge noise can be considered. When recover it, turn off the power and turn

on the power again for re-start of operation. When such takes place often times,
take an appropriate action for noise.

(3) No calibration available
<Cause and Treatment>

・Is the concentration of calibration gas correct ?

Prepare the appropriate calibration gas.

・There is the possibility for decreasing of sensor sensitivity.

The sensor replacement is required.

(4) PW/TR light flashing.
① System Trouble “E-00”
<Cause and Treatment>

・Memory trouble inside of detector.

Contact our nearest agent or Riken Keiki.

② Sensor connection trouble “E-01”
<Cause and Treatment>

・Non-connection of sensor, contact failure of connector and disconnection of sensor cable.

Check

“whether sensor is connected” or “the sensor cable connector is connected to the

sensor or amplifier board.

”

If can not be recovered, contact our nearest agent or Riken Keiki.

③ Abnormal function for zero correction 「E-04」

・Zero correction was out of available range due to sensor drift.

Contact our nearest agent or RIKEN KEIKI.

④ Low flow trouble

“E-05”

<Cause and Treatment>

・Drop of pump function

Even though the ball of flowmonitor can not be up the red line(lower red line) with FLOW
ADJ/▲▼switch, replace the pump with new one.

・Check that the detector unit is mounted firmly or the fixing screw of detector unit is

fastened firmly.