RKI Instruments GD-K71D User Manual
Page 30

PT2E-1052
29
5−5−4.Alarm test
This section is used to confirm the transmission condition to the outer equipment by giving
same gas concentration signal(gas concentration data) output.
①Press SET switch for above 3 seconds
to start alarm test mode. When it gets
to alarm test mode, both SKIP light and
display flash.
②Increase the reading with FLOW ADJ/▲▼
switch. When it reaches the first alarm
level, AL1 light turns on.
When it reaches the second alarm level,
AL2 light turns on.
● ● ○ ◎
PW/TR ALM1 ALM2 SKIP
When make alarm test, (transmission test) announce it to respective department
beforehand.
Carry it out after making proper treatment(External signal output, alarm contact).
Alarm contact does not work during alarm test even if the reading
reaches alarm level.
To confirm the operation of alarm contact, additional setting is required.
Please contact with our nearest agent or RIKEN KEIKI.
● ○ ○ ◎
PW/TR ALM1 ALM2 SKIP
WARNING
* Note
● ● ● ◎
PW/TR ALM1 ALM2 SKIP