5−5−2.Zero adjustment

This section is used to adjust zero level.






①Press MODE switch for above 3 seconds,

and the maintenance mode starts.

(SKIP light flashes)

②Select “1-1” by FLOW ADJ./▲▼switch

and press TEST/SET switch to start zero
adjusting mode. The zero suppression
circuit is released and actual zero level is
displayed.

③Supply fresh air. Press TEST/SET switch

after the reading gets stable, and the zero
is adjusted. (LED flashes during adjustment).








④After completion of zero adjustment,

press MODE switch to return the Menu.


⑤When return to detection mode, press MODE

switch again for above 3 sec.
The SKIP light turns off to show detection
mode









● ○ ○ ◎

PW/TR ALM1 ALM2 SKIP

● ○ ○ ◎

PW/TR ALM1 ALM2 SKIP

● ○ ○ ◎

PW/TR ALM1 ALM2 SKIP

● ○ ○ ○

PW/TR ALM1 ALM2 SKIP


After completion of adjustment, be sure to return detection mode(Only
PW/TR light is ON) by pressing MODE switch.

WARNING

CAUTION

* Note

When make zero adjustment, supply fresh air to the detector head and adjust it
after reading gets stable.

*Note

The reading may slightly changes after zero adjustment.
In such case, make fine adjustment with FLOW ADJ./
▲ ▼ switch. After that, press. TEST/SET switch.
Pressing TEST/SET switch, zero set is completed.



When leave it alone on maintenance mode, this returns to detection
mode automatically after 10 hours.