Maintenance check – RKI Instruments GD-K71D User Manual

Page 34

Advertising
background image

PT2E-1052

33

7.MAINTENANCE CHECK

This is an important for security and safety. To maintain the security and enhance
the reliability of safety, the regular maintenance check for it shall be absolutely
necessary.


7−1.Frequency of maintenance and check items.
7−1−1.Daily check

This is the check items carried out by customer.

①Check of flowmonitor

Check that the ball in flowmonitor of this unit is within two red lines.

②Check of PW/TR light.

In normal operation, the light is at on condition.

③Check of LED reading

Check that LED reading is zero “0”.
If it is not “0”, check that gas is free around gas sample point and adjust zero position

in maintenance mode.

④Check and replacement of dust filter
While this unit is working, ball in flowmonitor shall be positioned within two red lines. If it is

below the lower part of red line, adjust it with FLOW ADJ/▲▼switch to ▲.

Even though it cannot move up, the filth of dust filter can be considered. Replace it with
new one or check the flow path(clog, bend or crack of tubing).


7−1−2.Regular maintenance check

The following items shall be checked at regular check.

①Daily check
②Cleaning of this unit
③Adjustments
④Function check
⑤Parts replacement(Refer to 7-6 recommendable spare parts list)
⑥Others


7−1−3.Maintenance contract for regular check

To maintain the safety operation of the unit, it is recommended to keep the maintenance
contract with service agent for regular maintenance, adjustment and overhaul etc including
the gas sensitivity adjustment.
For the detail of maintenance contract, contact our nearest service agent or Riken Keiki.