Example on using table – Rockwell Automation 1440-VAD02-01RA XM-123 Aeroderivative Module User Manual

Page 153

Advertising
background image

Publication GMSI10-UM003D-EN-P - May 2010

145

Example on Using Table

The following example shows you how to use the Full Scale table to determine
the optimal Full Scale value.

To determine the optimal Full Scale value, follow these steps.

1. Refer to Table D.2 XM-123 Measuring Velocity with Accelerometer

(100 mV/g) on page 144.

2. Under the Maximum Vibration Level column, select the row that

corresponds to 10 ips.

3. Under the 10 Hz High Pass Filter (HPF) column, find the

recommended Full Scale Setting. The recommended Full Scale Setting
for the 10 Hz High Pass Filter is 0.17.

4. Refer to the value under the Max High Frequency Peak Amplitude

column to verify that there are not any signals present at the sensor that
exceed this value. For the example above, the Max High Frequency Peak
Amplitude value is 12 g’s.

If there are signals in excess of this level then increase the Full Scale
value to the next higher value, 0.5 in this example. If there are no

Table D.5 XM-123 Measuring Displacement with Displacement Sensor (200 mV/mil)

Max Vibration Level

Full Scale Setting

mils pp

micrometers pp

1Hz HPF

5Hz HPF

10Hz HPF

40Hz HPF

1000Hz HPF

5

125

1.5

N/A

50

1250

10

EXAMPLE

Application: XM-123 with 100 mV/g accelerometer

Units used for monitoring: velocity, ips

High Pass Filter: 10 Hz

Maximum vibration level: 5 ips pk (360 mm/s RMS)

TIP

Since the maximum vibration level of 5 ips is greater
than 3 ips, it is necessary to refer to the next higher
level in the table, which in this case is 10 ips.

Advertising