2 Installation

819 IC Detector / 820 IC Separation Center

24

2 Connect 819 to PC

•

Use 6.2125.110 Cable to connect Connection 9 “RS 232” of

the 819 IC Detector to a free RS 232 interface at the PC

(see Fig. 14).

3 Connect 819 and 820 to 771

•

Use 6.2134.140 Cable to connect Connection 5 “Output” of

the 819 IC Detector to Connection “Channel 1” of the 771

IC Compact Interface (see Fig. 14). The cable end with the

marking “Analog” is inserted into the detector.

•

Attach the two cable ends “RUN” and “COM” of 6.2134.140

Cable to the positions Integr. and Start of terminal block 27

of the 820 IC Separation Center (see Fig. 14).

3 Connect 771 to PC

•

Use 6.2134.100 Cable to connect Connection “RS 232, PC”

of the 771 IC Compact Interface to a free RS 232 interface

at the PC (see Fig. 14).

4 Connect column heating to PC

For 820.02X0 only:

•

Use 6.2134.040 Cable to connect RS 232 interface 22 of the

820 IC Separation Center to a free RS 232 interface at the

PC (see Fig. 14).


Fig. 14:

Connection of 819 and 820 to a PC

If a 2-channel system is used then the cable connections for the sec-

ond detector and second valve are made in a similar way.

771

819

820

6.2134.140

6.2125.110

6.2125.090

6.2134.040

6.2134.100