4 Notes - Maintenance - Faults

819 IC Detector / 820 IC Separation Center

78

No feed of regen-
eration or rinsing
solution for sup-
pressor

•

Contamination of the
filter in the 6.2821.120
Filter unit PEEK

•

Counterpressure too
high in suppressor
module

•

Replace the 6.2821.130
filter(see Section 2.7.3)

•

Clean or replace suppres-
sor module (see Section
4.2.6

…

4.2.8)

Detector is not
recognized by «IC
Net»

•

No RS connection

•

Switch instrument on

•

Check wiring

•

Check that the connection
specified in the 819 IC
Detector
window (see
Section 3.1.2 Links)
matches the actual wiring