2.4 Mains

connection

Follow the instructions below for connecting to the power supply. If

the instrument is operated with a mains voltage set wrongly and/or

wrong mains fuse, there is a danger of fire!

2.4.1

819 IC Detector

Setting the mains voltage

Before switching on the 819 IC Detector for the first time, check whether

the mains voltage set on the instrument (see Fig. 15) matches the local

mains voltage. If this is not the case, you must reset the mains voltage

on the instrument as follows:

1 Disconnect mains cable

Disconnect mains cable from mains connection plug 3 of the

819 IC Detector.

2 Remove fuse holder

Using a screwdriver, loosen fuse holder 4 below the mains con-

nection plug 3 and take out completely.

3 Change and check fuse if necessary

Carefully take the fuse installed for the desired mains voltage out

of fuse holder 4 and check its specifications (the position of the

fuse in the fuse holder is marked by the white arrow imprinted

next to the mains voltage range):

100

120 V 0.63 A (slow blow) Metrohm-No.

U.600.0014

220

240 V 0.315 A (slow blow) Metrohm-No. U.600.0011

4 Insert fuse

Change fuse if necessary and reinsert in fuse holder 4.

5 Install fuse holder

Depending on the desired mains voltage, insert fuse holder 4 in

the 819 IC Detector so that the corresponding mains voltage

range can be read normally and the adjacent white arrow points

to the white bar imprinted below the fuse holder (see Fig. 15).