2.9 Separating columns and suppressor module

819 IC Detector / 820 IC Separation Center

45

5 Connect suppressor module to detector block

(Suppressor connection 1)

Cut outlet capillary 69 (marked with “Detector”) of suppressor

module 70 to the required length. Use the 6.2126.080 Capil-

lary cutter available as an option.

Screw outlet capillary 69 on to coupling 71 by using a

6.2744.010 Pressure screw.

Screw inlet capillary 56 of detector block 57 on to the other

end of coupling 71.

6 Prepare 833 IC Liquid Handling Pump Unit

Mount the two lengths of 6.1826.050 Pump tubing (white-

yellow stopper) with aspirating tube 16 and filter unit PEEK

22 6.2821.120 as described in Section 3.2, Fig. 7 of the 833

Instructions for Use.

For mounting of the filter unit PEEK see also section 2.7.3.

Pump tubing is a consumable whose working life depends on the

contact pressure. This is why when the pump is switched off for a long

time, the tubing cassette should be lifted by fully loosening spring

lever on the right-hand side (in this way the correctly set optimal con-

tact pressure is retained).

7 Suppressor connection 2: H

2

SO

4

Pull inlet capillary 65 (marked with "H

2

SO

4

") by hand as far as

required through one of the feedthroughs 14 to the outside

(see Fig. 23 und Fig. 24).

Screw the filter unit PEEK on to PEEK coupling 21

(6.2744.160) as described in Section 3.2 of 833 Instructions

for Use at the outlet end of the first piece of pump tubing 19.

Fasten inlet capillary 65 (23) to the outlet of filter unit PEEK

(6.2821.120) 47 by using a compression fitting screw (see

Fig. 7 of 833 Instructions for Use).

At the inlet end of the first piece of tubing 19 fasten a suitably

long piece of PTFE tubing 16 (6.1803.020) to coupling 18

(6.2744.034) by using a compression fitting (6.2744.010) (see

Fig. 7 of 833 Instructions for Use).

Immerse the other end of the aspiration tubing in a container

with regeneration solution (normally 50 mmol/L H

2

SO

4

) and fix

it in position.

Pull outlet capillary 66 of the suppressor module (marked

with "Waste") through opening 18 on the rear panel, lead it

into a sufficiently large waste bottle and fix it in position.