Metrohm 820 IC Separation Center User Manual

Page 40

Advertising
background image

2 Installation

819 IC Detector / 820 IC Separation Center

32

3 Installation of the capillary in the IC Separation

Center

•

Unscrew the four knurled screws 15 of the bottom rear panel

28 of the 820 IC Separation Center and remove rear panel

(see Fig. 5).

•

Insert PEEK capillary 48 from the back into the inner com-

partment of the IC Separation Center.

•

Install rear panel 28 so that the capillary is positioned in

opening 34 „Inlet A“ or 32 „Inlet B“ and screw on with the four

knurled screws 15.

4 Connection of the pulsation dampener

•

Procedure see section 2.7.2.

5 Connection to the injection valve

•

Connect PEEK capillary 48 to pulsation dampener 49 (see

Section 2.7.2). Using another PEEK capillary 48, connect pul-

sation dampener to connection „5“ (see Fig. 28) of injection

valve 51.

6 Mount column connection capillary

This step is only necessary, if no column heating is used.

•

Connect column connection capillary50 (ca. 20 cm of

6.1831.010 PEEK capillary) to connection „4“ (see Fig. 28) of

injection valve 51.