4 Notes - Maintenance - Faults

819 IC Detector / 820 IC Separation Center

74

6 Clean connection piece

•

Clean the sealing surface of connection piece 73 using a lint-

free cloth and ethanol.


Fig. 29:

Assembling the suppressor

72 Screw nut

73 Connection piece with input and

output leads

(6.2832.010)

for suppressor module

74 Suppressor rotor(6.2832.000)

75 Suppressor holder

7 Insert connection piece

•

Insert connection piece 73 in suppressor holder 75 in such a

way that connection "1" is at the top, and that the three lugs

on the connection piece fit in the corresponding openings of

the holder.

•

Screw nut 72 onto the thread of suppressor holder 75 manu-

ally (do not use tools).

8 Connect and condition the suppressor

•

Reconnect the suppressor to the IC system.

•

Before switching the suppressor to the next position for the

first time, rinse all 3 suppressor units with solution for 5 min.