4 Notes - Maintenance - Faults

819 IC Detector / 820 IC Separation Center

76

6 Insert connection piece

•

Insert connection piece 73 in suppressor holder 75 in such a

way that connection "1" is at the top, and that the three lugs

on the connection piece fit in the corresponding openings of

the holder.

•

Screw nut 72 onto the thread of suppressor holder 75 manu-

ally (do not use tools).

7 Connect and condition the suppressor

•

Reconnect the suppressor to the IC system.

•

Before switching the suppressor to the next position for the

first time, rinse all 3 suppressor units with solution for 5 min.