8 precolumns, Precolumns – Metrohm 820 IC Separation Center User Manual

Page 41

Advertising
background image

2.8 Precolumns

819 IC Detector / 820 IC Separation Center

33

2.8 Precolumns

The use of easily exchangeable precolumns protects the separating

columns and prolongs their lifetime. The precolumns available from

Metrohm are either real precolumns or precolumn cartridges, which are

used together with a cartridge holder. For the installation of a precol-

umn cartridge into the accompanying cartridge holder see the attached

leaflet.

The precolumn that is suitable for your separating column can be

found in the

Metrohm IC Column Catalog which can be obtained

from your local Metrohm agency, the data sheet accompanying your

separating column, the product information about separating columns

that can be found under

http://www.metrohm.com

, ion chromatogra-

phy products, or let your agency advise you directly.

New IC precolumns are normally filled with solution and sealed at

both ends. Before the precolumn is installed in the system, it must be

ensured that this solution is freely miscible with the eluent used

(check manufacturer's specifications).

1 Connect precolumn

Remove end caps from the precolumn.

Fit compression fitting to the connection capillary 50

mounted to the injection valve (see Section 2.6).

Screw column connection capillary 50 to precolumn.

Cut off a small piece, ca. 5 cm, from the PEEK capillary

6.1831.010 and fit compression fittings to both ends (see

Section 2.6).

Mount the prepared capillary to the outlet of the precolumn.

When you install the column, always ensure that this is inserted cor-

rectly in accordance with the flow direction (if existent) shown.

2 Rinse the precolumn

Place a beaker beneath outlet capillary.

Switch on 818 IC Pump in «IC Net» and rinse precolumn for

ca. 10 min with eluent.

Switch off 818 IC Pump.