2 Installation

819 IC Detector / 820 IC Separation Center

44

3 Connect column to suppressor module (Suppressor

connection 1)

•

Cut inlet capillary 64 (marked with "Eluent") of suppressor

module 70 to the required length. Use the 6.2126.080 Capil-

lary cutter available as an option.

•

without column heating

Screw inlet capillary 64 on to the outlet end of separating col-

umn 53 using a 6.2744.010 Compression fitting.

•

with column heating

Connect capillary 52 at the outlet of the separating column

(see Fig. 20) to the inlet capillary 64 using the 6.2744.040

Coupling.

4 Fix column

•

without column heating

Insert one or two column holders 59 (6.2027.030, 6.2027.040

or 6.2027.050) in the mounting rails 58 and fasten separating

column in the column holder.

•

with column heating

Insert column heating according to Fig. 10 into the Separation

Center.

Fig. 24:

Suppressor module connections


64 Suppressor inlet capillary for

eluent

65 Suppressor inlet capillary for

H

2

SO

4

66 Suppressor outlet capillary

for H

2

SO

4

67 Suppressor outlet capillary

for H

2

O

68 Suppressor inlet capillary for

H

2

O

69 Suppressor outlet capillary

for eluent

Eluent

H

2

O

Detector

Waste

Waste

H

2

SO

4

1

3

2

65

64

69

66

68

67