6.6 Subwindow Titrants/Solutions

■■■■■■■■■■■■■■■■■■■■■■

1384

■■■■■■■■

tiamo 2.3 Patch 1

Input

24 characters

Default value

'empty'

Serial number

Serial number of dosing unit. With intelligent dosing units the serial num-
ber is read off automatically and cannot be edited.

Input

10 characters

Default value

'empty'

Cylinder volume

Cylinder volume of dosing unit. With intelligent dosing units the cylinder
volume is read off automatically and cannot be edited. If in the method
you have selected a solution in a titration or dosing command then the
cylinder volume will be checked in the run.

Selection

2 mL | 5 mL | 10 mL | 20 mL | 50 mL

Default value

20 mL

Cylinder serial number

Serial number of cylinder. It is printed on new cylinders. With intelligent
dosing units it is read off automatically. It can be altered at any time, e.g.
if the cylinder has to be replaced.

Input

8 characters

Default value

'empty'

Parameters for preparation
Configures the parameters to be used in the commands PREP and
EMPTY.

Dosing port Prep/Empty

Dosing port via which the cylinder contents are to be ejected during prep-
aration and emptying.

Selection

Dosing port 1 | Dosing port 2 | Fill port | Spe-
cial port

Default value

Dosing port 1

Dosing rate Dosing port 1

Speed at which dosing is to be carried out at Dosing port 1. The maxi-
mum dosing rate depends on the cylinder volume of the exchange unit
used (see below). When the function is carried out the speed will auto-
matically be reduced to the largest possible value.

Range

0.01 ... 166.00 mL/min

Selection

maximum

Default value

maximum