5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

952

■■■■■■■■

tiamo 2.3 Patch 1

continuous
A temperature sensor must be connected. The temperature will then
be measured continuously.
automatic
If a temperature sensor is connected then the temperature will be
measured continuously. Otherwise, the temperature entered manually
under Temperature on the Measuring parameters tab will be
used.
off
The temperature will not be measured. The temperature entered man-
ually under Temperature on the Measuring parameters tab will be
used.

Stirrer

Stirrer

Selection of the stirrer.

Titrando, 855, 867
Selection

1 | 2 | 3 | 4 | off

off
Means that no stirrer will be used.

Stirring rate

Setting the stirring rate. The algebraic sign of the stirring rate changes the
direction in which the stirring is done.

Titrando, 855, 867
Range

-15 ... 15

Default value

8

Switch off automatically

on | off (Default value: on)

If this option is activated, the stirrer will be switched off automatically
when the command has finished. This parameter is displayed only for
Titrandos, 855 and 867.

5.6.4.4.3

MEAS U - Measuring parameters

Tab: Method

MEAS U Properties... Measuring parameters

Command name

Name of the command.

Input

25 characters

The parameters for the measuring procedure can be defined on this tab.