2 met cond - general/hardware, General/hardware – Metrohm tiamo 2.3 Patch 1 User Manual

Page 693

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.3 Patch 1

■■■■■■■■

677

Designation

Description

.SVT

Total start volume (volume that was added in
dependence on all three start conditions) in mL

.TC.TC

Temperature coefficient in %/°C

.TC.CO

Coefficient c0 of the Chebyshev polynomial of the
solution used

.TC.C1

Coefficient c1 of the Chebyshev polynomial of the
solution used

.TC.C2

Coefficient c2 of the Chebyshev polynomial of the
solution used

.TC.C3

Coefficient c3 of the Chebyshev polynomial of the
solution used

.TC.C4

Coefficient c4 of the Chebyshev polynomial of the
solution used

.TC.MAX

Maximum temperature coefficient in %/°C

.TC.MIN

Minimum temperature coefficient in %/°C

.TC.TSTART

Start temperature in °C

.TC.TSTOP

Stop temperature in °C

5.6.3.3.6.2

MET Cond - General/Hardware

Tab: Method

MET Cond Properties… General/Hardware

Command name

Name of the command.

Input

25 characters

The general parameters for the control device, the dosing device, the sen-
sor and the stirrer are defined on this tab.

Device

Device name

Selection of a device from those available in the device table. The only
devices which shall be offered are those with which the command can
run.

Selection

'Device name' | not defined

Default value

not defined

not defined
In this case, the device name must be assigned by the user at the start
of the method.

Advertising