5 Method

tiamo 2.3 Patch 1

■■■■■■■■

833

Range

0 ... 999999 s

Default value

0 s

5.6.3.6.4

KFC - Control parameters

Tab: Method

KFC Properties... Control parameters

Command name

Name of the command.

Input

25 characters

Endpoint

EP with

Measured value for the endpoint.

756, 831
Range

-2000.0 ... 2000.0 mV

Default value

50.0 mV

851, 852
Range

-1250.0 ... 1250.0 mV

Default value

50.0 mV

Titration rate

Titration rate

The three predefined sets of parameters can be selected for the titration
rate: slow, optimal and fast; the parameters for Control are not dis-
played. The User setting must be selected in order to be able to edit these
parameters.

Selection

slow | optimal | fast | User

Default value

optimal

slow
For titrations in which the finest details are also to be visible. This could
however also lead to an increase in noise, which could result in unwan-
ted EPs.
optimal
Parameter set for all standard titrations; optimized for the most fre-
quent applications.
fast
For less critical, fast titrations.
User
User-defined settings of the individual titration parameters which affect
the titration rate.