3.8 Dosing and liquid handling
Metrohm Sample Processor, Operation
107
Formation of separating bubble (air gap)
11
Command sequence: 
 
LIFT: 1 : shift mm
DOS: 1.1 : -0.5 mL
In order to prevent the carrier liquid and 
sample solution from mixing a separating 
bubble, must be formed that occupies a 
length of at least 5 mm in the pipetting 
tubing. However, the separating bubble 
must also be measured out with sufficient 
accuracy. This is why a Dosing unit with 
max. 20 mL cylinder capacity should be 
used for pipetting. 
Move to sample
Command sequence:
MOVE: 1 : sample
LIFT: 1 : work mm
The (dissolved or liquid) samples can be
contained in open beakers on the rack or 
in sealed vials. In the latter case instead 
of the pipetting tubing a suitable injec-
tion needle should be connected to Dos-
ino Port 1 with a normal piece of tubing 
(2 mm dia.). 
 
Aspirate sample
11
Command sequence:
DOS: 1.1 : -5.0 mL
WAIT: pause 3 s
LIFT: 1 : shift mm
DOS: 1.1 : -0.2 mL
Aspiration of the sample should be car-
ried out at a reduced filling rate 
(<10 mL/min). The appropriate setting 
can be made under <PARAM>
>Dosing
unit def.
With non-aqueous samples an additional 
small air bubble may be drawn into the 
pipetting tip in order to prevent the sam-
ple solution from dripping. 
Move to target
222
Before the pipetting tubing
containing the measured 
out sample is moved, the 
corresponding Dosino port 
must be closed. 
 
Command sequence: 
 
DOS: 1.2 : port mL
MOVE 1 : ext.1
LIFT: 1 : work mm
The titration cell should 
be filled with solvent 
before the sample is 
added so that the
sample can be pipet-
ted directly into the liq-
uid (tip immersed).