3.8 Dosing and liquid handling
108
Metrohm Sample Processor, Operation
Eject sample
11
Command sequence:
DOS: 1.1 : EndVol mL
The sample should be ejected at a re-
duced dosing rate (<10 mL/min). The 
appropriate settings can be made under 
<PARAM> 
>Dosing unit def.
Rinse
11
Command sequence: 
 
DOS: 1.2 : fill mL
DOS: 1.1 : endVol mL
DOS: 1.1 : compen. mL
or
DOS: 1.2 : -x.xx mL
DOS: 1.1 : endVol mL
DOS: 1.1 : compen. mL
In order to keep sample solution carry-
over as low as possible, rinsing with sol-
vent can be carried out after the sample 
has been ejected. This is particularly rec-
ommended for non-aqueous samples. 
In preparation for the next pipetting proc-
ess the cylinder should be completely 
ejected. 