4.3 Error messages

Metrohm Sample Processor, Service, maintenance, errors

121

∗ invalid rack code

The rack code read in by the Sample Processor cannot be

found in the internal table.

∗ missing beaker

After a MOVE command no beaker can be recognized at the

selected position.

∗ rack data missing

No sample rack in position, or no rack data can be found for

the sample rack currently on the turntable.

∗ RS232 error

The transmission parameters of the RS232 interface do not

coincide with those of the receiving device.

∗ SCAN timeout

The connected device has not transmitted the expected sig-

nal within the defined timeout time. The sample determina-

tion may not have been carried out properly, or the connec-

tion has been interrupted. Check the connected device.

* service recommended

The warning limit of the running time meter has been

reached. It is time for the Sample Processor to be serviced.

Please contact your local Metrohm service department.

∗ user memory full

The memory for the user-defined methods is full. Before new

methods can be stored methods that are not used, or only

used infrequently, must be deleted.

∗ wrong rack

The rack placed on the turntable is not the rack assigned to

the method under

parameters

.

trap error xxx

Unexpected program error; switch instrument off and then on

again.

No display, LEDs Tower 1

and Tower 2 light up

LCD error (system error 7). Contact Service.