3 rack definitions – Metrohm 789 Robotic Sample Processor XL User Manual

Page 69

Advertising
background image

3.2

Configuration

Metrohm Sample Processor, Operation

61

3.2.3 Rack

definitions

configuration

>rack definitions

Submenu for defining individual racks

Open the submenu with <ENTER>

>rack definitions

>>recall rack

Loads rack definitions

In order to make a modification to the definition of a sample rack the rack

data must first be loaded. The data of the Metrohm standard racks is stored

under their ordering number.

>recall rack

name: 6.2041.310

Name of the rack to be loaded

10 ASCII characters

The <SELECT> key can be used to make a selection from the saved rack
data. The rack data is loaded with <ENTER>. The first selection to appear
is the rack name of the currently attached rack.

>recall rack

code 000001

Magnet code of the rack
See table on p. 91.

000001…111111


The magnet code is used for the unambiguous identification of the rack. The

magnet code is recognized during the initialization of the rack. This is why
the <RACK> key should be pressed whenever a rack is changed.

>recall rack

work position T1

0 mm

Working position for sample positions at

Tower 1 *

)

0…235 mm

(in mm from upper stop point)

>recall rack

work position T2

0 mm

Working position for sample positions at

Tower 2 *

)

0…235 mm

(in mm from upper stop point)

>recall rack

rinse position T1

0 mm

Rinsing position for sample positions at
Tower 1 *

)

0…235 mm

(in mm from upper stop point)

>recall rack

rinse position T2

0 mm

Rinsing position for sample positions at
Tower 2 *

)

0…235 mm

(in mm from upper stop point)


>recall rack

shift position T1

0 mm

Shifting position for sample positions at
Tower 1 *

)

0…235 mm

(in mm from upper stop point)

Use <QUIT>

for next higher

level