Metrohm 789 Robotic Sample Processor XL User Manual

Page 27

Advertising
background image

1.7 The keypad

Metrohm Sample Processor, Introduction

19

fined binary pattern (<SELECT> selection), e.g.

START device 1

,

INIT

etc.


2nd parameter, for RS232 interface

Data string with up to 14 alphanumeric characters (any).
The default value "&M;$G" (for starting Metrohm instruments) can be
set with <CLEAR>.

3

WAIT

The <WAIT> key has no function in the normal operating condition. It
is used to insert the WAIT command in a run sequence.

0

DEF

The <DEF> key is used to edit various settings for manual opera-

tion. Repeated pressing of the <DEF> key is used to select the vari-
ous settings. In order to change the entry you must first press
<ENTER> and then enter the new value.

Alterations made in this way only apply to manual operation.

DEF settings

0

DEF

DOSRATE

Change dosing rate

• The dosing rate in mL/min can be set separately for each dosing

device, see p. 70.

• Syntax:

DOSRATE

[dosing device] [dosing rate]

0

DEF

FILLRATE


Change filling rate


• The filling rate in mL/min can be set separately for each dosing

device, see p. 70.

• Syntax:

FILLRATE

[dosing device] [filling rate]

0

DEF

COCKMOVE


Direction of stopcock rotation

• For each connected Dosino the direction of rotation of the stop-

cock switching can be defined separately, see p. 70.

• Syntax:

COCKMOVE

[Dosing device] [Direction of rotation]

0

DEF

LIFTRATE


Change lift rate

• The lift speed in mm/s can be set separately for each tower (for 2-

tower models), see p. 68.

• Syntax:

LIFTRATE

[Tower] [Lift rate]